• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Current repair: product evaluation product review of the air dermatome serial number (b)(4) by (b)(4) on 30 april 2020 revealed that the rpms were out of specification, the control bar was not in the correct position, the calibration was out at the 0 reading.The motor, pins, and planetary carrier were replaced to get the rpms in specification.Product repair: repair of the device was performed by (b)(4) on 7 may 2020 which included replacement of the following: motor, pins, planetary carrier, bearings, seal.Additional repair included recalibration the device, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.Previous repair: air dermatome serial number (b)(4) has not been previously repaired/evaluated before 30 april 2020 as documented in the repair reports in livelink.Review of the device history records identified no deviations or anomalies during manufacturing.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the unit was sent in for preventive maintenance and found in need of repair.Device was sent in for preventive maintenance only.No adverse event was reported as a result of this malfunction.Upon investigation it was found that the motor was running below motor speed specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key10267908
MDR Text Key198743883
Report Number0001526350-2020-00594
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375901
UDI-Public(01)00889024375901
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number63415170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-