• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Catalog Number BXA077901J
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem Occlusion (1984)
Event Date 02/13/2020
Event Type  Injury  
Event Description
The following was reported to gore: on (b)(6) 2020, this patient underwent endovascular treatment using three gore® viabahn® vbx balloon expandable endoprosthesis (vbx) for leriche syndrome.Two vbx were deployed in the aortic bifurcation using the kissing stent technique.A third vbx was deployed to extend the left device distally.The patient tolerated the procedure.On (b)(6) 2020, it was reported the left and right vbx had collapsed.The left vbx device was relined using a bare metal stent (brand unknown).Plain old balloon angioplasty (poba) was performed to treat the collapse of the right vbx device.The patient tolerated the procedure.
 
Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Engineering evaluation: the referenced complaint did not have a returned device.The following evaluation is based on information obtained from the event description and communication summary.Crush resistance of the vbx device can be influenced by the stent ring material, stent ring geometry, stent ring spacing, and stent ring alignment.The stent rings are certified by the supplier for the material and geometric properties.Stent rings are 100% verified during the manufacturing process to be free of visual deformities.Stent ring geometry, spacing, and alignment are 100% verified during the manufacturing process.The device history file was reviewed and no anomalies were identified.Based on the device evaluation performed, no manufacturing anomalies were identified to which the event could be definitively attributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Type of Device
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
MDR Report Key10271709
MDR Text Key198752036
Report Number2017233-2020-01028
Device Sequence Number1
Product Code PRL
Combination Product (y/n)Y
PMA/PMN Number
P160021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2022
Device Catalogue NumberBXA077901J
Was Device Available for Evaluation? No
Date Manufacturer Received01/14/2021
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
Patient Age87 YR
-
-