Model Number CI-1600-04 |
Device Problems
Mechanical Problem (1384); No Audible Prompt/Feedback (2282)
|
Patient Problem
Therapeutic Response, Decreased (2271)
|
Event Date 09/24/2020 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing decreased performance and sound quality issues.Programming adjustments were made, however, the issue did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient is reportedly experiencing intermittent tinnitus.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual inspection.The photographic imaging inspection revealed severed electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The device passed the electrical tests performed.The scanning electron microscopy (sem) analysis revealed corrosion on an electrode.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|