MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP Back to Search Results
Catalog Number A2114
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 10/21/2019
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Between (b)(6) 2019 and (b)(6) 2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from (b)(6).Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa (b)(4) have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period (b)(6) 2019 through (b)(6) 2020.Integra lifesciences contacted (b)(6), director of regulatory programs, office of product evaluation and quality and (b)(6), assistant director, mdr team, office of product evaluation and quality on (b)(6) 2020 to report these issues regarding mdr reports.
 
Event Description
A customer sent the a2114 mayfield infinity xr2 skull clamp for maintenance.Integra service and repair supervisor found the left and right pawls broken and no longer functioning properly when the product was received from the customer.The device was not used in a surgery thus there was no patient contact nor injury to the patient.There was no surgery delay.Additional information was been received on 11nov2019 stating the incident date was on (b)(6) 2019.The product problem was discovered before surgery and they used another device after the product problem occurred.
 
Manufacturer Narrative
Additional information - d10, g4, g7, h2, h3, h6, h10 device identifier # (b)(4).The device was returned for evaluation.Unit cleaned per protocol.Unit received with the left and right pawls broken and needed to be replaced.Unit received without pedi rocker arm or suit case.General maintenance and cleaning required.The device exceeded its expected life of 7 years.The reported complaint was confirmed from the evaluation of item.The complaint is likely caused by improper handling.The definite root cause cannot be reliably determined.Device identifier #: (b)(4).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 SKULL CLAMP
Type of Device
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10276194
MDR Text Key199672590
Report Number3004608878-2019-01111
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K090506
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2114
Device Lot Number2009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2019
Initial Date Manufacturer Received 11/05/2019
Initial Date FDA Received07/15/2020
Supplement Dates Manufacturer Received02/04/2020
Supplement Dates FDA Received07/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-