ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information obtained during product evaluation.Current repair product evaluation product review of the air dermatome serial number (b)(4) by dover on (b)(6) 2020 revealed that the motor was noisy, calibration out at the 0,20,10 readings, head has visible wear, and the width plates need to be replaced.The device is over 10 years old.Product repair of the device was performed by dover on (b)(6) 2020 which included replacement of the following: head, motor, bearings, gears, width plates, spring seal, o-rings, reciprocating arm, eccentric shaft, fine adjustment cams, thickness control lever, and various other parts additional repair included recalibration the device, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.Previous repair air dermatome serial number (b)(4) has not been previously repaired/evaluated before (b)(6) 2020 as documented in the repair reports in livelink.The device history record (dhr) review was unable to be performed as the dhr associated with this device is not available.This device was manufactured prior to may 2009 where it was identified a robust dhr indexing and handling process was not in place.An action implemented a serial number logbook to correct this issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the unit was sent in for preventative maintenance and found in need of repair.Device was sent in for preventative maintenance only.No adverse events were reported.Upon inspection- width plates were worn/damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key10276604
MDR Text Key200025797
Report Number0001526350-2020-00606
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-