Updated fields: d4, d10, g4, g7, h2, h3, h4, h6, h10 the perforator was received for evaluation device history record - based on the dhr review conducted, there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Device identifier # (b)(4).Failure analysis - visual inspection showed no anomalies.Results of instructions for use, unit was found to performed as intended.Results of functional test was found to performed as intended the root cause is undetermined and was unable to be confirmed in the complaint evaluation.
|