RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/09/2020
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the codman disposable perforator 14mm seized up and stopped drilling bone half way through skull while performing a stereotactic biopsy.
 
Manufacturer Narrative
Updated fields: d4, d10, g4, g7, h2, h3, h4, h6, h10 the perforator was received for evaluation device history record - based on the dhr review conducted, there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Device identifier # (b)(4).Failure analysis - visual inspection showed no anomalies.Results of instructions for use, unit was found to performed as intended.Results of functional test was found to performed as intended the root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10280212
MDR Text Key203541830
Report Number1226348-2020-00210
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 04/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2020
Date Manufacturer Received07/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-