MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA STANDARD INFINITY SKULL CLAMP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA STANDARD INFINITY SKULL CLAMP; N/A Back to Search Results
Catalog Number A1114A
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A customer reported that the a1114a standard infinity skull clamp would not slide in and out or was stuck.There was no know patient injury or delay in surgery.
 
Manufacturer Narrative
Additional information: d4, udi#: (b)(4).The device was returned to manufacturer for evaluation.Complaint was confirmed, as the ratchet extension arm would not go through the base smoothly, and the unit had multiple worn out components that caused the unit to fail multiple specific functional tests.The device was manufactured in 2007.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD INFINITY SKULL CLAMP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10280257
MDR Text Key203320962
Report Number3004608878-2020-00197
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K051440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1114A
Device Lot Number079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2020
Initial Date Manufacturer Received 04/02/2020
Initial Date FDA Received07/15/2020
Supplement Dates Manufacturer Received06/02/2020
Supplement Dates FDA Received07/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-