Catalog Number 71930020
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2020
Event Type  malfunction  
Event Description
It was reported that during a thr surgery was noticed that the tip of the corkscrew fem head remover is no longer sharp and is bent.The procedure was finished using a smith and nephew back up device, with no surgical delay and no injury to the patient.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirms the tip of the device is dull and bent.The device shows significant signs of wear/usage.The device was manufactured in 2016.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORKSCREW FEM HEAD REMOVER
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10280407
MDR Text Key199043464
Report Number1020279-2020-03244
Device Sequence Number1
Product Code HRY
UDI-Device Identifier00885556049709
UDI-Public885556049709
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71930020
Device Lot Number16LM07650
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-