• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP Back to Search Results
Catalog Number A2114
Device Problems Break (1069); Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A customer reported that the rocker arm of the a2114 mayfield infinity xr2 skull clamp was stuck and broken.There was no known patient injury nor delay in surgery.
 
Manufacturer Narrative
Additional information - d10, g4, g7, h2, h3, h4, h6, h10.Udi # (b)(4).The device was returned for evaluation.The device history record (dhr) documentation showed no abnormalities related to the reported failure.The reported complaint was confirmed as the unit was received with a broken rocker arm assembly in two pieces.Physical damage or breakage noted on inspection.Most likely due to excessive force applied to arm.Definite root cause cannot be determined at this time.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 SKULL CLAMP
Type of Device
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10280843
MDR Text Key203191028
Report Number3004608878-2020-00300
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K130389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2020
Date Manufacturer Received06/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-