MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Injury (2348)
Event Date 05/22/2020
Event Type  Injury  
Manufacturer Narrative
Udi: (b)(4).Perforator was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.
 
Event Description
A physician reported a minor dural injury during the use of the perforator, and dural repair (suture) was performed.The procedure was completed with the reported product and there was over 30 minutes of surgical delay.Procedure: clipping of cerebral aneurysm.
 
Event Description
N/a.
 
Manufacturer Narrative
(b)(4).The perforator was returned for evaluation: device history record - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye: some organic matter was present.Complaint could not be verified.The perforator was found to meet all acceptance criteria.Unit was found to perform as intended (ifu test, functional test) the root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10280924
MDR Text Key202696852
Report Number1226348-2020-00304
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ61C34
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Initial Date Manufacturer Received 05/26/2020
Initial Date FDA Received07/15/2020
Supplement Dates Manufacturer Received08/07/2020
Supplement Dates FDA Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-