Unique device identifier (udi)- (b)(4).Additional information received on 08 july 2020, stating that the medical device gets stuck in the skull, needing the help of a pliers and the male force to remove it.A tear of the dura is generated.No treatment was done; over time the tear caused closed.Another craniotomy kit was used to complete the procedure and guarantee patient care.Perforator was not returned for evaluation therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|