• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Mechanical Problem (1384)
Patient Problem Injury (2348)
Event Date 05/11/2020
Event Type  Injury  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the perforator was blocked and impacted the skull.The specialist needed a plier to retrieve the device from the skull.Patient is recovering.There was a few minutes surgical delay due to product problem.
 
Event Description
N/a.
 
Manufacturer Narrative
(b)(4).Perforator was returned for evaluation: device history record (dhr)- there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis- the perforator unit was inspected using the unaided eye.No anomalies were observed.Ifu testing procedure was performed with no observed anomalies.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.The returned unit was found to work as intended, and met all acceptance criteria.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Manufacturer Narrative
Unique device identifier (udi)- (b)(4).Additional information received on 08 july 2020, stating that the medical device gets stuck in the skull, needing the help of a pliers and the male force to remove it.A tear of the dura is generated.No treatment was done; over time the tear caused closed.Another craniotomy kit was used to complete the procedure and guarantee patient care.Perforator was not returned for evaluation therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10280946
MDR Text Key202597583
Report Number1226348-2020-00316
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup,Followup
Report Date 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ57C77
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2020
Date Manufacturer Received09/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age26 YR
-
-