• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML Back to Search Results
Model Number CSE-P-125-JA
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/08/2020
Event Type  malfunction  
Manufacturer Narrative
No patient information was provided at this time.The bowl from the cell saver elite set was returned and evaluated by haemonetics.It was noted during investigation that blood was seen on the inner core.Bowl decompression testing confirmed the blood outflow from the inner core and the outer core welding and a crack was confirmed in the inner core of bowl.
 
Event Description
On (b)(6) 2020 haemonetics was notified of a long empty error message which was displayed on the 5th cycle of the return process during a cardiac procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 125ml.A total blood volume of 3800ml was processed and there was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer Contact
janice kiser
125 summer street
boston, ma 
MDR Report Key10283105
MDR Text Key202559237
Report Number1219343-2020-00066
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/04/2022
Device Model NumberCSE-P-125-JA
Device Lot Number0120047
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/17/2020
Initial Date FDA Received07/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-