MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the a1059 mayfield modified skull clamp, the ratcheting mechanism was sticking.It is unknown if there was any patient contact or injury.Additional information has been sent.
 
Manufacturer Narrative
The device was returned for evaluation.No dhr review was possible as the provided serial number (b)(6) does not appear to be valid integra identification number.The unit was received having both rotational and lateral movement and a buildup of residue.Upon disassembly, it was noted that the index knob and lock needed new components to replace worn internal parts.Index knob is old and needs to be updated.Unit needs to be machined to have heli-coils added to the large starburst threads.The set screw in the swivel base was tight.The reported complaint was confirmed via inspection of the unit.The lock is difficult to turn and sticking due to heavy residue build up in the locking mechanism.Udi # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10283158
MDR Text Key203675520
Report Number3004608878-2020-00120
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,user facility
Type of Report Initial,Followup
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2020
Date Manufacturer Received04/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-