Catalog Number A1059 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported that the a1059 mayfield modified skull clamp, the ratcheting mechanism was sticking.It is unknown if there was any patient contact or injury.Additional information has been sent.
|
|
Manufacturer Narrative
|
The device was returned for evaluation.No dhr review was possible as the provided serial number (b)(6) does not appear to be valid integra identification number.The unit was received having both rotational and lateral movement and a buildup of residue.Upon disassembly, it was noted that the index knob and lock needed new components to replace worn internal parts.Index knob is old and needs to be updated.Unit needs to be machined to have heli-coils added to the large starburst threads.The set screw in the swivel base was tight.The reported complaint was confirmed via inspection of the unit.The lock is difficult to turn and sticking due to heavy residue build up in the locking mechanism.Udi # (b)(4).
|
|