CLARET MEDICAL, INC. SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE Back to Search Results
Model Number CMS15-10C-US
Device Problems Positioning Failure (1158); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/25/2020
Event Type  malfunction  
Event Description
It was reported that the proximal filter was unable to be resheathed.The patient's anatomy contained a medium level of tortuosity and a medium level of calcium.A sentinel embolic protection system was advanced over a choice pt extra support wire.The proximal filter of the sentinel embolic protection system was able to be deployed.The physician was not able to deploy the distal filter of the sentinel embolic protection system.At the end of the procedure, the proximal filter was not able to be fully resheathed before removal from the patient.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
CLARET MEDICAL, INC.
1745 copperhill parkway
santa rosa CA 95403
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
160 knowles drive
los gatos CA 95032
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10284770
MDR Text Key199136499
Report Number2134265-2020-09450
Device Sequence Number1
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public00863229000004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/26/2022
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number0025293855
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-