Freedom driver s/n (b)(6) passed all testing requirements defined per (b)(4), freedom driver system final test and acceptance procedure during incoming inspection.Visual inspection of driver exterior found no damage or abnormalities.Visual inspection of driver interior identified particulate debris on top of the piston cylinder assembly (pca) and primary and secondary motors.The secondary motor was also found out of default position indicating secondary motor engagement.Alarm history data review found three new alarms recorded in the driver; two "2d" fault code alarms which occur as a result of secondary motor engagement and one "4a" fault code alarm, which can occur when the driver is left running while disconnected from a patient or patient simulator.Testing was performed to replicate the asterisks displayed for fill volume and cardiac output and permanent alarm.Manual manipulation of the airflow sensor cable produced fill volume and cardiac output display reading fluctuations as well as an alarm.An extended observation run of 48 hours did not produce any additional alarms or display errors.The root cause of the permanent fault alarm is most likely attributed to the patient condition, i.E., coughing.Excessive coughing can cause a sudden increase in the pulmonary pressure which will cause a momentary drop in cardiac output and can cause the driver to switch to the secondary motor.Since evidence of secondary motor engagement was found during incoming inspection, this is the most likely cause of the permanent fault alarm.The root cause of the asterisks displayed for cardiac output and fill volume as reported by the customer and display reading fluctuations observed during testing is most likely an intermittent connection in the cable that connects the airflow sensor to the main pcba.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.Ce (b)(4).Follow-up report 1.
|