SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 72200616 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/06/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during a shoulder surgery, the motor drive unit was not working.The procedure was completed without delay using a back-up device.No patient injury or other complications were reported.Preliminary results of investigation have concluded that this unit displayed a blade stall error which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Event Description
|
It was reported that during a shoulder surgery, the motor drive unit was not working.The procedure was completed without delay using a back-up device.No patient injury or other complications were reported.Results of investigation have concluded that this unit displayed a blade stall error which makes it a reportable event.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The housing and the gearbox were removed from the motor.The gearbox was found to be jammed.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A blade stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.
|
|
Search Alerts/Recalls
|
|
|