Model Number G247
Device Problem Inaccurate Synchronization (1609)
Patient Problem Ventricular Tachycardia (2132)
Event Date 03/11/2020
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that the patient was in the emergency room and upon interrogation was noted to have received anti-tachycardia pacing (atp) due to ventricular tachycardia (vt) from the cardiac resynchronization therapy defibrillator (crt-d).Review of the episode found that there was a ventricular pace that was near the t wave right before the episode that may have induced the vt.One burst of atp was given and the rate fell just below the cut off rate at the end of charging thus the shock was diverted.The crt-d remains implanted and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key10287311
MDR Text Key199227334
Report Number2124215-2020-07721
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/15/2020
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number188634
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/08/2020
Initial Date FDA Received07/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age62 YR
-
-