Udi # (b)(4).The device was returned for evaluation.The device history record (dhr) showed no abnormalities related to the reported failure.The provided lot code 119 indicates a 2011 manufacture date.This device is over 7 years old.Integra¿s recommended product life is 7 years based upon usage as listed in the dhf.A review of previous a1059 complaints related to "lock position" found that upon evaluation of many of these devices, the locks had movement, a residue buildup was present, and worn internal components of the lock required replacement.Given the age of this device, it is possible the reported issues are due to wear and lack of maintenance.However, without a return device, the complaint cannot be confirmed, and the root cause cannot be reliably determined.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|