ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 07/02/2020
Event Type  malfunction  
Manufacturer Narrative
During preventive maintenance, the autopulse platform (sn: (b)(4)) displayed "system error, out of service, revert to manual cpr" error message upon powering on.The root cause for the observed issue was due to a communication error between the drivetrain motor and the processor board, as a result of normal wear and tear.The autopulse platform was manufactured in october 2005 and is almost 15 years old, well beyond the expected service life of 5 years.The platform was connected to the autopulse vision software, and the error message was cleared.There was no physical damage observed on the returned autopulse platform during the visual inspection.The archive data review showed occurrence of system error, user advisory (ua) 136 (internal parameter corrupted) error message.Service will not be performed as the customer declined repair.The autopulse platform will be returned to the customer without the repair, and it will have a label state "not for clinical use".Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
The autopulse platform (sn: (b)(4)) was returned for preventive maintenance (pm).During the functional testing, the autopulse platform displayed "system error, out of service, revert to manual cpr" error message upon powering on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10292982
MDR Text Key199666479
Report Number3010617000-2020-00727
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000130
UDI-Public00849111000130
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-