• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-401L-0523
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The needle could not be extended from the outer tube.Injection could not be performed while the slider was pushed, but injection could be performed while the slider was pulled.The outer tube was buckled at 280 mm, 395 mm and 560mm from the distal end.The needle tube was compressive buckled.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the frictional resistance between the outer tube and the needle tube increased due to the bending of the tube or pushing the slider fast, and then the needle tube was compressive buckled when the needle was extended and injection could not be performed.The above device handling has warned in the instruction manual as follows.Straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
 
Event Description
During an unspecified procedure, the subject device was used.The needle could not be extended.The intended procedure was completed with another device.There was no patient injury reported.On july 3, 2020, olympus medical systems corp.(omsc) found that the subject device could not perform the injection.This is the report regarding the failure of injection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Type of Device
INJECTOR AND SHEATHSET
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10294120
MDR Text Key223497572
Report Number8010047-2020-04586
Device Sequence Number1
Product Code FBK
UDI-Device Identifier04953170382352
UDI-Public04953170382352
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K902736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberNM-401L-0523
Device Lot NumberK9X23
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/03/2020
Initial Date FDA Received07/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-