Model Number 1217-25-500 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Adhesion(s) (1695); Pain (1994); No Code Available (3191)
|
Event Date 07/09/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the patient presents with groin pain.Surgeon believes cup could be loose.Head is removed, followed by liner, screw and hole eliminator.Cup insertion handle is attached to shell and found to be well fixed and appropriate orientation.Range of motion did show some impingement with internal rotation.Trial reduction was most stable with a +4/10 degree liner and one longer ball.Scar tissue was removed from where the impingement was taking place.Doi: (b)(6) 2017; dor: (b)(6) 2020; right hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected: d11.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary :no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
|