• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Product Quality Problem (1506); Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/07/2020
Event Type  malfunction  
Manufacturer Narrative
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that during the procedure, the physician was unable to pass the guidewire through the tip of the left ventricular lead.The inner lumen was damaged upon inspection.The physician suspected it was due to packaging or the guidewire.Another lead was used.There was no patient consequences.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key10303415
MDR Text Key199729646
Report Number2938836-2020-07539
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000094694
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/07/2020
Initial Date FDA Received07/21/2020
Supplement Dates Manufacturer Received08/10/2020
Supplement Dates FDA Received08/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-