MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 06/26/2020
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of "the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message" was confirmed during the functional testing and on the archive data review.The root cause for ua7 was due to defective load cells.The cracked front enclosure and defective load cell were likely attributed to mishandling such as a drop.Upon visual inspection, cracked front enclosure was observed likely due to user mishandling such as a drop.The initial functional testing of the autopulse platform could not be performed due to ua 07 error message displayed upon powering on.The archive data review showed occurrence of multiple ua07 error message.The load sensing system has detected a weight/load imbalance between the two load cells.Both load cells needs to be replaced to address the ua07 error.In addition, unrelated to the reported complaint, noticed that the date and time in the archive data are corrupted.The processor board needs to be replaced to address the date and time issue.The probable root cause for the observed date and time issue is due to normal wear and tear.The autopulse platform was manufactured in 2007 and is 12 years old, well beyond the expected service life of 5 years.Waiting for customer's approval for service.Historical complaints were reviewed for service information related to the reported complaint and there was one similar complaint reported for autopulse with serial number (b)(4).Ccr (b)(4) reported on 02 nov 2011, both load cells were replaced.
 
Event Description
During shift check, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message on the user control panel.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10304876
MDR Text Key199893549
Report Number3010617000-2020-00697
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/29/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-