The reported complaint of "the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message" was confirmed during the functional testing and on the archive data review.The root cause for ua7 was due to defective load cells.The cracked front enclosure and defective load cell were likely attributed to mishandling such as a drop.Upon visual inspection, cracked front enclosure was observed likely due to user mishandling such as a drop.The initial functional testing of the autopulse platform could not be performed due to ua 07 error message displayed upon powering on.The archive data review showed occurrence of multiple ua07 error message.The load sensing system has detected a weight/load imbalance between the two load cells.Both load cells needs to be replaced to address the ua07 error.In addition, unrelated to the reported complaint, noticed that the date and time in the archive data are corrupted.The processor board needs to be replaced to address the date and time issue.The probable root cause for the observed date and time issue is due to normal wear and tear.The autopulse platform was manufactured in 2007 and is 12 years old, well beyond the expected service life of 5 years.Waiting for customer's approval for service.Historical complaints were reviewed for service information related to the reported complaint and there was one similar complaint reported for autopulse with serial number (b)(4).Ccr (b)(4) reported on 02 nov 2011, both load cells were replaced.
|