DEPUY INTERNATIONAL LTD - 8010379 UNKNOWN HIP FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD
Device Problem Naturally Worn (2988)
Patient Problems Pain (1994); Test Result (2695); No Code Available (3191)
Event Date 11/22/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
The literature article entitled, "capsular closure outweighs head size in preventing dislocation following revision total hip arthroplasty" written by christopher pedneault, dylan tanzer, anas nooh, karen smith and michael tanser published by hip international accepted by publisher on 22 november 2018 was reviewed.The purpose of the article was to compare the dislocation rates in revision cases that received either 28 mm, 32 mm or large 36 mm femoral head components.Data was compiled from 48 hips, 47 hips and 49 hips respectively.Article reports 3 of the 144 hips dislocated.It is noted that 2 of the 3 hips were identified with depuy products and captured in this complaint as follows: first case is 32 mm head group a (b)(6) years old male who underwent acetabular liner exchange for excessive poly wear and pelvic osteolysis.The acetabular component 56 mm duraloc cup and srom stem were well fixed and left in place.A 10 degree oblique poly liner oriented superiorly was inserted to compensate for the 57 degree abduction angle of the cup.The 28 mm head was replaced with a 32 mm +3 co cr head.The patient then experienced a posterior dislocation 11 months postop and treated with non-operative posterior hip precautions for 3 months and suffered no further dislocations at 2 years follow up.Depuy product and adverse event: poly liner - revised for wear and osteolysis; duraloc cup - mispositioned but left in situ; femoral head and poly liner - dislocation treated non-operatively.Third case is (b)(6) years old male with failed metal-metal-metal asr xl tha presenting with pain, severe acetabular osteolysis, elevated metal ions but no pseudotumor.Intraoperatively, depuy trilock stem was well fixed and left in situ, and the acetabular and femoral components were replaced with non-depuy components which was followed by a dislocation.He was treated non-operatively with posterior hip precautions and suffered no further dislocations.Depuy product and adverse event: asr xl cup, asr xl head, revised for pain, acetabular osteolysis, elevated metal ions (attributed to bearing wear).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key10306313
MDR Text Key200795157
Report Number1818910-2020-16486
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Date Manufacturer Received08/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-