MSD DEGGENDORF MFG CLYDESDALE PTC SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
Model Number 4986855 |
Device Problem
Break (1069)
|
Patient Problems
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
Event Date 07/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from healthcare provider via manufacturing representative regarding patient with degenerative scoliosis for spinal therapy.Event occurrred intra-op.It was reported the intervertebral space was narrow, the trial was not possible due to the influence of the iliac bone, and when tried to insert the cage without fully applying distraction, the cage broke at the inserter gripping part during insertion.It was said that it seemed that the doctor had hit it strongly when inserting the cage.It was said that the cage was broken into pieces when it was taken out, and about half of it remained in the body.There was delay in the procedure of less than 60 min.No patient complications or symptoms were reported.No further complication reported.Update (b)(6) 2020; it was planned to perform olif at l2-5.The procedure started from l4/5, and the cage broke.
|
|
Manufacturer Narrative
|
H3: product analysis #(b)(4):4986855.Lot# h5574075 a visual review of the returned implant identified that portions have been broken.Microscopic examination of the inserter interface identified damage to the threaded hole area and deformation on the top face, and fracture on one side around the inserter tang interface, consistent with significant impact and torsional loading.Microscopic examination of the fracture surface of the broken portions of the implant identified a fracture with rays emanating away from the area of crack initiation, consistent with brittle overload.The observations are consistent with impaction overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from healthcare provider via manufacturing representative regarding patient with degenerative scoliosis for spinal therapy.Event occurred intra-op.It was reported the intervertebral space was narrow, the trial was not possible due to the influence of the iliac bone, and when tried to insert the cage without fully applying distraction, the cage broke at the inserter gripping part during insertion.It was said that it seemed that the doctor had hit it strongly when inserting the cage.It was said that the cage was broken into pieces when it was taken out, and about half of it remained in the body.There was delay in the procedure of less than 60 min.No patient complications or symptoms were reported.No further complication reported.Update (b)(6) 2020; it was planned to perform olif at l2-5.The procedure started from l4/5, and the cage broke.
|
|