• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550350-23
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Patient Involvement (2645)
Event Date 06/04/2020
Event Type  malfunction  
Manufacturer Narrative
Device location is unknown as account could not be determined/reached.Investigation is not yet complete.A follow-up report will be submitted with all relevant information.
 
Event Description
User facility medwatch report received that states stent was not on balloon and properly secured to the stent delivery system.Rt caught the error prior to using on patient.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10309019
MDR Text Key199911395
Report Number2024168-2020-06104
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648227431
UDI-Public08717648227431
Combination Product (y/n)Y
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/19/2021
Device Model Number1550350-23
Device Catalogue Number1550350-23
Device Lot Number0011041
Was Device Available for Evaluation? No
Date Manufacturer Received07/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-