• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74121142
Device Problems Insufficient Information (3190); Patient Device Interaction Problem (4001)
Patient Problem Injury (2348)
Event Date 01/11/2011
Event Type  Injury  
Event Description
It was reported that a right hip revision surgery was performed due to unknown reasons on (b)(6) 2011.The new implants are from a competitors brand.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the bhr head was removed.The cup remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr head and cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the cup.Similar complaints have been identified for the head.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.Based on the information provided, the reported pain and moderate tissue reaction may be consistent with findings associated with metal debris, although intraoperatively, the acetabular component was noted to be secure.Based on the provided information the source of the reported findings cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and expected post-operative healing/pain cannot be determined.No further clinical assessment is warranted at this time.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 42MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key10310648
MDR Text Key199984228
Report Number3005975929-2020-00255
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502773
UDI-Public03596010502773
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 08/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number74121142
Device Catalogue Number74121142
Device Lot Number080848
Initial Date Manufacturer Received 06/26/2020
Initial Date FDA Received07/22/2020
Supplement Dates Manufacturer Received08/28/2020
Supplement Dates FDA Received08/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-