Udi number: (b)(4).The device was not returned to edwards for evaluation as it remains implanted.If additional information is received a supplemental mdr will be submitted.The device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Peri-procedural arrhythmias, heart block, and other conduction disturbances are common in patients with underlying cardiovascular disease and/or conduction abnormalities.They can be exacerbated with standard perioperative medications, anesthesia, and/or instrumentation of the heart.Temporary pacemakers are inserted in all patients undergoing aortic valve replacement (avr).It is not uncommon for patients to have short term/reversible periods of heart block or arrhythmias following the procedure while their heart recovers from cardiopulmonary bypass.In many cases, the temporary pm is left in the patient for a short time following the procedure and then subsequently removed prior to discharge.Following surgical aortic valve replacement (avr), new-onset bundle branch block has been reported in 16% to 32% of patients and the need for permanent pacemakers in 3% to 8% of patients.The reason for post-operative av block after surgical avr is related to injury to the cardiac conduction system during surgical excision of the adjacent diseased valve and annular tissue.The close anatomical relationship between the aortic valvular complex and the branching atrioventricular bundle explains the possible development of conduction abnormalities following prosthetic aortic valve procedures.Arrhythmias and conduction disturbances are also common after surgical repair or replacement of the tricuspid and mitral valves due to the proximity of the conduction pathways.The cause of the event cannot be determined; however, surgical technique and patient factors may have contributed to the event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.It was learned patient implanted with a 21mm aortic valve had dual-chamber pacemaker placed on post-operative day five due to complete heart block.The patient was discharged to inpatient rehab on post-operative day 12 and discharged home on post-operative day 15.
|