Catalog Number A-6000-08LF
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual nor functional inspection of the product involved in the complaint could not be conducted since the product was not returned.The device history record of batch number 74m1900321 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.No corrective action can be established at this moment since the product sample is not available for evaluation.Product sample is necessary to perform a proper investigation to determine the root cause and the corresponding corrective actions.The customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility neither is being manufactured at the time.If defective sample becomes available at a later date this complaint will be updated as applicable.Teleflex will continue to monitor and trend related events.
 
Event Description
The report states: they installed the thoracic drainage system, all the drainage came out from the bottom where the red and blue valve is located, it tried to adjust and it was not possible that it will stop draining.Therefore, a system change must be made.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED DRY/ WET LF
Type of Device
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key10315956
MDR Text Key200816095
Report Number3004365956-2020-00141
Device Sequence Number1
Product Code KDQ
Combination Product (y/n)N
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Type of Report Initial
Report Date 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/08/2022
Device Catalogue NumberA-6000-08LF
Device Lot Number74M1900321
Was Device Available for Evaluation? No
Date Manufacturer Received06/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-