Catalog Number A1114
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was not yet received by the manufacturer for evaluation.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report number 3004608878-2020-00398, 3004608878-2020-00399, and 3004608878-2020-00401.
 
Event Description
This is 3 of 4 reports.A customer reported that the locking knob of the a1114 mayfield infinity skull clamp was too loose.There was no known patient injury or surgery delay.
 
Event Description
N/a.
 
Manufacturer Narrative
(b)(4).The device was returned for evaluation.The lock has rotational and lateral movement and a residue buildup was present.The unit needed new components added to replace worn internal parts.The device history record (dhr) showed no abnormalities related to the reported failure.The reported complaint was confirmed.The observed condition was likely caused by wear and tear of the device; the definite root cause could not be reliably determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10319230
MDR Text Key200255841
Report Number3004608878-2020-00400
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K051440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Date Manufacturer Received07/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-