Model Number 11500A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Arrhythmia (1721); Complete Heart Block (2627)
Event Date 06/26/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned to edwards for evaluation as it remains implanted.The device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Peri-procedural arrhythmias, heart block, and other conduction disturbances are common in patients with underlying cardiovascular disease and/or conduction abnormalities.They can be exacerbated with standard perioperative medications, anesthesia, and/or instrumentation of the heart.Temporary pacemakers are inserted in all patients undergoing aortic valve replacement (avr).It is not uncommon for patients to have short term/reversible periods of heart block or arrhythmias following the procedure while their heart recovers from cardiopulmonary bypass.In many cases, the temporary pm is left in the patient for a short time following the procedure and then subsequently removed prior to discharge.Following surgical aortic valve replacement (avr), new-onset bundle branch block has been reported in 16% to 32% of patients and the need for permanent pacemakers in 3% to 8% of patients.The reason for post-operative av block after surgical avr is related to injury to the cardiac conduction system during surgical excision of the adjacent diseased valve and annular tissue.The close anatomical relationship between the aortic valvular complex and the branching atrioventricular bundle explains the possible development of conduction abnormalities following prosthetic aortic valve procedures.Arrhythmias and conduction disturbances are also common after surgical repair or replacement of the tricuspid and mitral valves due to the proximity of the conduction pathways.In this case, the physician could not exclude a relationship with the device.The patient received a permanent pacemaker.A manufacturing related issue was not identified.A definitive root cause could not be determined.If additional information is received a supplemental mdr will be submitted.Edwards lifesciences will continue to monitor all reported events.No further actions are required at this time.
 
Event Description
Edwards received information that a patient experienced an arrhythmia event soon after the implantation of a 25mm valve.As reported, a 3rd degree atrioventricular block occurred on pod # 3.A permanent pacemaker was implanted and the outcome was noted as to be resolved.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRIS RESILIA AORTIC VALVE
Type of Device
TISSUE, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key10320218
MDR Text Key200420299
Report Number2015691-2020-12668
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,s
Type of Report Initial,Followup
Report Date 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/16/2022
Device Model Number11500A
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-