Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 06/22/2020
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of "the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message" was confirmed during the functional testing and on the archive data review.The root cause for the reported complaint was due to the defective load cell2, likely due to a defective component or harsh impact due to user handling.Visual inspection was performed and found no physical damage to the autopulse platform.The initial functional testing of the autopulse platform could not be performed due to user advisory (ua) 07 error message displayed upon powering on, thus confirming the customer complaint.Review of the archive data indicated error message user advisory (ua) 07 on the customer reported event date.The load cell characterization test revealed that the load cell2 was over-reporting.The defective load cell needs to be replaced to address the error.Waiting for customer approval for repair.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with serial number (b)(4).
 
Event Description
During the training, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message.The error could not be cleared by the user.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, ca 
MDR Report Key10324143
MDR Text Key200433029
Report Number3010617000-2020-00718
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-