MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems No Display/Image (1183); Visual Prompts will not Clear (2281)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of "the customer was unable to read the displayed user advisory because the lcd screen was unreadable" was confirmed during the functional testing.The root cause of the reported complaint was a defective backlight of the lcd display, likely as a result of normal wear and tear.The autopulse platform is a reusable device and was manufactured in 2010, and it is 10 years old, well beyond its expected service life of 5 years.The reported complaint of "the user was unable to clear the unknown displayed user advisory" was confirmed based on the archive data review.Per archive, the last time the autopulse platform was used to compress an object similar to a human was on 09 july 2020, and there was a user advisory (ua) 20 (position out of range) error message observed to have occurred on that date.Based on the archive, the user was unable to clear the error message, and the use of the autopulse platform was discontinued; thus, confirming the reported complaint.The ua20 was not replicated during the functional testing, and no device malfunction was found that could have caused or contributed to the occurrence of the ua20 error message.The probable root cause of the ua20 error was the driveshaft not to be at "home" position, most likely attributed to unintended user error.User advisory is the clearable error message and is designed into the platform to alert the operator that autopulse has detected one of several conditions.Per auto pulse user advisory list, user advisory 20 alerts when the drive shaft is not within the specified range of positions.This typically occurs if the autopulse attempted to perform compressions with a cut lifeband installed which allowed the driveshaft to rotate out of operating range.Also, this error message can be triggered due to an internal component error or malfunction.This error message can be cleared by returning the drive shaft to home position using the administrative menu.Visual inspection of the returned platform was performed, and no physical damage was observed.The autopulse platform passed the initial functional testing without any fault or error.During the functional testing, it was noted that the backlight of the lcd was not lighting up; however, the display was still readable in daylight circumstances.The defective lcd display was replaced to remedy the fault.Following service, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The brake gap inspection was performed and verified the brake gap was within the specification.Load cell characterization test was performed and confirmed both cell modules are functioning within the specification.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
During patient use, the customer was unable to read the user advisory (ua) that was displayed by the autopulse platform (sn: (b)(4)) because the lcd screen was unreadable.The user was unable to clear the unknown displayed user advisory, and therefore, the use of the autopulse platform was discontinued, and manual cpr was performed.No further information was provided by the customer.Patient's status information was requested, but the customer did not provide a response; therefore, patient's status is unknown.As reported, the autopulse platform backlight was noted to be defective a few months prior to the reported event; however, the customer had declined to send the device for service/repair because the lcd screen was still readable at the time.Per the archive data review, the unknown error message observed by the customer was user advisory (ua) 20 (position out of range) occurred on july 09, 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10324367
MDR Text Key200433605
Report Number3010617000-2020-00753
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001076
UDI-Public00849111001076
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0740-16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-