It was reported that a revision surgery was performed on the patients left hip due to right hip pain and right leg weakness, failure of hip resurfacing, markedly elevated metal levels, intramuscular lesion in posterior right buttock.The patient outcome is unknown.
|
It was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr head.Similar complaints have been identified for the cup.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.It was reported that a revision surgery was performed on the patients left hip due to right hip pain and right leg weakness, failure of hip resurfacing, markedly elevated metal levels, intramuscular lesion in posterior right buttock.Smith and nephew has not received the device/adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.Currently, it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|