MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090W
Device Problems Over-Sensing (1438); Connection Problem (2900); Output Problem (3005); Data Problem (3196)
Patient Problem No Patient Involvement (2645)
Event Date 07/01/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that oversensing was noted on the analyzer atrial channel of the programmer.60 cycle noise was noted on the electro cardiogram (ecg) cable.The cables were swapped and but the noise persisted.The noise could be eliminated by pushing sideways on the ecg connector port with the noise returning when pressure was released.The programmer was returned for service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis confirmed the customer's comment as the electrocardiogram (ecg) signal was noisy due to a loose ecg connector.It was also noted that the stylus tip was loose but functional.One hinge plate screw was missing.The link electronic module (lem) printed circuit board was replaced and calibrated.The stylus was replaced and the overlay calibrated.Installed and secured hinge plate screw.Hard drive was reconfigured and software updated and reloaded.Circuit boards and mechanical parts were inspected.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key10326099
MDR Text Key200414771
Report Number2182208-2020-01360
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00885074301440
UDI-Public00885074301440
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090W
Device Catalogue Number2090W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Date Manufacturer Received07/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
2067 RADIOFREQUENCY HEAD, 229047 ANALYZER.; 2067 RADIOFREQUENCY HEAD, 229047 ANALYZER
-
-