The device was returned without the packaging.The reported lot number was not confirmed.During visual inspection, the stent was returned deployed and was intact.The stent delivery wire was kinked bent.Functional inspection was not carried out as stent was deployed.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was returned for analysis and the stent was seen to be deployed and the sdw was kinked bent.The as reported event will be assigned procedural factors as this complaint appears to be associated with a product that met stryker and design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.
|
It was reported that during the procedure, the stent (subject device) had difficulty inserting inside the hub of a microcatheter.After the subject stent was inserted successfully, it had difficulty advancing inside the microcatheter.The physician made multiple attempts to advance the subject stent but the delivery sheath came off and the subject stent deployed prematurely inside the catheter shaft.The physician used a snare device to remove the subject stent from the patient¿s anatomy.The subject stent was replaced and the procedure was completed successfully.There were no clinical consequences to the patient.
|