• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS Back to Search Results
Model Number A22041A
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/24/2020
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic hysterectomy procedure, the ceramic insulation at the distal end of the resection sheath broke off and fell into the patient.However, no fragment remained inside the patient since it was reportedly retrieved.The intended procedure was successfully completed with the same set of equipment and there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation.The type of damage described by the user is typically caused by thermal and/or mechanical overload in combination with wear and tear.Therefore, this event/incident was attributed to use error.It cannot be conclusively determined whether the insulating insert had already been pre-damaged before the incident occurred, whether the damage was triggered during the reprocessing cycle preceding the incident, or during the last procedure.Furthermore, it is reported that the sealing ring is damaged.Sealing rings are wear parts and must be replaced on a regular basis.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath without showing any abnormalities.The case will be closed on olympus side and the user will be informed about the investigation results.Furthermore, but, independently from the above mentioned issue, a capa was initiated in march 2019 where further investigations, trending, and surveillance will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 24 FR.
Type of Device
RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key10331747
MDR Text Key200679253
Report Number9610773-2020-00168
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020961
UDI-Public04042761020961
Combination Product (y/n)N
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22041A
Device Catalogue NumberA22041A
Device Lot Number121W
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received07/28/2020
Supplement Dates Manufacturer Received08/05/2020
Supplement Dates FDA Received08/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-