• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ENDURON 10D 66OD X 28ID; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ENDURON 10D 66OD X 28ID; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 124166025
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); No Code Available (3191)
Event Date 07/16/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient's hip was revised due to dislocation.The primary femoral head and duraloc liner were replaced by a larger diameter head and a duraloc constrained liner.Doi: (b)(6) 2004.Dor: (b)(6) 2020; affected side: left hip.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.¿.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary : examination of the returned acetabular liner finds nothing outward to suggest product error.It is noted however that the device was badly damaged when extracted.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.H10 additional narrative:.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURON 10D 66OD X 28ID
Type of Device
DURALOC IMPLANT : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10332776
MDR Text Key200620273
Report Number1818910-2020-16975
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number124166025
Device Lot NumberT28FT1014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2020
Date Manufacturer Received09/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARTICUL/EZE BALL 28 +8.5 BL; ARTICUL/EZE BALL 28 +8.5 BL
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-