• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLT,PLASMA,ISBT SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLT,PLASMA,ISBT SET Back to Search Results
Model Number 80300
Device Problems Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Reaction (2414)
Event Date 02/22/2020
Event Type  Injury  
Manufacturer Narrative
Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported an allergic/hypotensive transfusion reaction during single dose platelet (sdp) transfusions on a patient with thrombocytopenia post chemotherapy due to cancer return.The clinician at the customer site ordered administration of antihistamines, steroids and beta blockers.Per the customer, the patient is normal.Patient gender is not available at this time.The disposable set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Upon review, no reported donor reactions have been reported on this trima lot number.The run data file (rdf) was analyzed for this event.There is no evidence or suspicion of device malfunction based on the rundata file analysis.According to the aabb circular of information for the use of human blood and blood components (revised 2013).Anaphylactic reactions characterized by hypotension, tachycardia, nausea, vomiting and/or diarrhea, abdominal pain, severe dyspnea, pulmonary and/or laryngeal edema, and bronchospasm and/or laryngospasm, are rare but dangerous complications requiring immediate treatment with epinephrine.These reactions have been reported in iga-deficient patients who develop antibodies to iga antibodies investigation is in process.A follow up report will be provided.
 
Manufacturer Narrative
Investigation: a disposable history search was performed for this lot.One additional report of a similar issue from the same customer was found.That event was reported on mdr 1722028- 2020-00350 and 1722028-2020-00351.Updated investigation: according to the aabb circular of information for the use of human blood and blood components (revised 2017).Anaphylactic reactions characterized by hypotension, tachycardia, nausea, vomiting and/or diarrhea, abdominal pain, severe dyspnea, pulmonary and/or laryngeal edema, and bronchospasm and/or laryngospasm, are rare (<10/100,000 transfused units) but dangerous complications requiring immediate treatment with epinephrine and supportive care.While these reactions have been reported in iga-deficient patients who develop antibodies to iga antibodies with anti-iga antibodies and patients with haptoglobin deficiency, most reactions are idiosyncratic and not associated with a specific serum protein deficiency, polymorphism, or identifiable cause.Root cause: there is no evidence or suspicion of device malfunction based on the run data file analysis.A definitive root cause could not be determined.Possible causes include but are not limited to: - patient's underlying other allergies.- patients with severe iga deficiency may develop antibody to iga and, with repeated transfusion, are at high risk of allergic reaction.- patients with haptoglobin deficiency may develop reactions that are idiosyncratic and not associated with a specific serum protein deficiency, polymorphism, or identifiable cause.- presence of donor and product attributes such as pro-allergic plasma ige or immunogenic proteins in the platelet product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL PLT,PLASMA,ISBT SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key10332885
MDR Text Key201922655
Report Number1722028-2020-00352
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583803008
UDI-Public05020583803008
Combination Product (y/n)N
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/01/2021
Device Model Number80300
Device Catalogue Number80300
Device Lot Number1911044151
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 07/07/2020
Initial Date FDA Received07/28/2020
Supplement Dates Manufacturer Received08/11/2020
09/08/2020
Supplement Dates FDA Received08/21/2020
09/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age00068 YR
Patient Weight70
-
-