Five samples were returned.Six samples were not returned.There were two cases with a method code of testing of actual/suspected device, a results code of operational problem identified, and a conclusion code of cause cannot be traced to device.There were six cases with a method code of device not returned, a results code of no findings available, and a conclusion code of cause not established.There was one case with a method code of testing of actual/suspected device, a results code of operational problem identified, and a conclusion code of cause not established.There was one case with a method code of testing of actual/suspected device, a results code of no device problem found, and a conclusion code of failure to follow instructions and cause cannot be traced to device.There was one case with a method code of testing of actual/suspected device, a results code of no device problem found, and a conclusion code of failure to follow instructions.The manufacturer internal reference number is: (b)(4).
|