One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Multiple short circuits were detected during electrical testing.Fluid was noted outside of the irrigation tubing but within the tygon tubing proximal to the catheter handle.In addition, dried saline was noted within the connector plug, consistent with the short circuits detected.Further testing revealed a leak at the luer/irrigation tubing junction due to a gap in the adhesive between the irrigation tube and the proximal tube inside the luer lock, allowing fluid to flow outside of the irrigation tubing and into the proximal tubing towards the electrical connector, consistent with the dried saline noted within connector plug.
|