Model Number N/A |
Device Problem
Failure to Cut (2587)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery the dermatome was running but did not cut through the skin.There was no harm or delay reported.Even though there was damage to the harvested skin, the graft harvest was able to be used.An alternate device was available for use to complete the procedure.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).D4 udi: (b)(4).Review of the most recent repair record determined the motor was slow and depth bar had side play.The motor was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
|
|
Event Description
|
No additional event information available.
|
|