The complaint component was returned and analyzed, and the reported allegation of failure to cycle was not confirmed via product analysis.The ambicor cylinders, pump and tubing were visually inspected.Both cylinders had multiple leaks in the rear tip cylinder body that was result of sharp instrument damage; holes were present.Cylinder 2 had broken fabric treads in cylinder body.There was no damage to the pump.The krt identified to be cut separated with sharp instrument damage.Based on the results of this investigation, no escalation is required.
|