Model Number CD3369-40Q |
Device Problems
Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
|
Patient Problem
Therapeutic Effects, Unexpected (2099)
|
Event Date 07/21/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that a patient presented remotely with a implantable cardioverter defibrillator (icd) that inappropriately shocked the patient for supra ventricular tachycardia (svt).Patient was scheduled to come in for programming changes but not done as of yet.No patient consequences reported.
|
|
Event Description
|
New information notes that programming changes were made to address the issue.The patient was stable.
|
|
Event Description
|
New information notes that patient received appropriate anti-tachycardia pacing for an arrhythmia.After the arrhythmia was converted, the device continued to deliver therapy inappropriately.Programming changes made to address this issue.Patient was stable.
|
|