MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 RECLAIM PROX NECK TRL 45MM; RECLAIM INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 RECLAIM PROX NECK TRL 45MM; RECLAIM INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Model Number 2975-31-045
Device Problems Loose or Intermittent Connection (1371); Device-Device Incompatibility (2919); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/23/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the reclaim trial neck and body are not functioning properly.The neck will spin on the body due to the body being worn.Also, the neck tend to get stuck in the locked position.These are the original trial necks that have since been redesigned.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Corrected: h6 (device) from device-device incompatibility to mechanical jam.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information states that the instrument did not break into two pieces.The grooves on the proximal body was the feature that was worn and not making it function.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the instrument associated with this report was not returned.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM PROX NECK TRL 45MM
Type of Device
RECLAIM INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10339771
MDR Text Key200876891
Report Number1818910-2020-17159
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295156376
UDI-Public10603295156376
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 07/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2975-31-045
Device Catalogue Number297531045
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/23/2020
Initial Date FDA Received07/29/2020
Supplement Dates Manufacturer Received08/04/2020
09/10/2020
09/17/2020
Supplement Dates FDA Received08/07/2020
09/15/2020
09/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
-
-