The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68) and a penumbra system aspiration pump max 220 (pump max).During the procedure, while the pump max was powered on, the physician felt some shaking when the distal end of the ace68 moved on the monitor and subsequently, the physician decided to remove the ace68.Upon removal of the ace68, it was noticed that the distal end of the ace68 was stretched.Therefore, the ace68 was not used for the remainder of the procedure.The procedure was completed using a new ace68 and the same pump max.There was no report of an adverse effect to the patient.
|
Results: the returned ace68 was kinked at approximately 128.0 cm, 129.0 cm, and 132.0 cm from the hub.The total length was measured approximately 133.0 cm from the hub.Conclusions: evaluation of the returned ace68 revealed kinks on its distal shaft.If the device is advanced against resistance, damage such as distal kinks may occur.During functional testing, water was aspirated through the ace68 and the reported shaking on the distal end was not observed.The total length of the ace68 was measured within specification; therefore, the reported distal end stretching could not be confirmed.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|