• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801763
Device Problem Communication or Transmission Problem (2896)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Per the field service representative (fsr), the end user stated that the small roller pump was not working and the user facility's manufacturer trained staff performed troubleshooting and determined that the nic board needed to be replaced.The fsr could not verify the reported issue.He replaced the nic board.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the network interface card (nic) board had a bad port.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.Per data log analysis, on 07-jul-2020 the power manager was reporting repetitive network interface card (nic) brick #3 state change from 'connected with fault' to 'connected and good'.This indicates a problem with the nic port or something connected to the nic port.The log confirms the complaint.During laboratory analysis, the product surveillance technician (pst) was not able to verify the reported issue.There were no functional issues observed with the nic panel during the evaluation.The system detects modules connected to all 12 ports and continues to operate reliably.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10347227
MDR Text Key201562131
Report Number1828100-2020-00281
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-