Model Number 801763 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Per the field service representative (fsr), the end user stated that the small roller pump was not working and the user facility's manufacturer trained staff performed troubleshooting and determined that the nic board needed to be replaced.The fsr could not verify the reported issue.He replaced the nic board.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the network interface card (nic) board had a bad port.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Per data log analysis, on 07-jul-2020 the power manager was reporting repetitive network interface card (nic) brick #3 state change from 'connected with fault' to 'connected and good'.This indicates a problem with the nic port or something connected to the nic port.The log confirms the complaint.During laboratory analysis, the product surveillance technician (pst) was not able to verify the reported issue.There were no functional issues observed with the nic panel during the evaluation.The system detects modules connected to all 12 ports and continues to operate reliably.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|