Model Number 1458Q/86
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 03/10/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2017865-2020-10277.It was reported that the patient complaint of pain at the implantable cardioverter-defibrillator (icd) site since implant.Further investigation noted the left ventricular (lv) lead had dislodged the day after implant, and every vector tested had phrenic nerve stimulation.The dislodgement was confirmed on x-ray the day after implant.The lv lead was turned off due to lack of lv options for revision.The patient was brought into lab for pocket revision and lv lead explant.The lv lead was explanted with no complications.The physician elected to explant and replace the icd as well.The patient was stable and was discharged.
 
Manufacturer Narrative
Additional information: d10, h3, h6, h10 as received, a complete lead was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies with the exception of procedural damage.The full measured s-curve hump height was within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10350224
MDR Text Key201206347
Report Number2017865-2020-10278
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000084131
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
QUADRA ASSURA MP ICD; QUADRA ASSURA MP ICD
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-