A physician reported that during an intraocular lens (iol) implant procedure, the room was too cold, estimated to be below 16 degrees as the thermometer had not been working.The iol was loaded having heated the viscoelastic (with a hand) and the iol in the heating cabinet.The iol was stiff and did not move easily, so was thought not safe to proceed with insertion.A backup lens was loaded very quickly and insertion proceeded smoothly after a heating cabinet was used to bring the iol, viscoelastic and instruments up to room temperature.
|
Product evaluation: the product was not returned for analysis.Additional information: the file states that "the temperature of the theatre was set at 14 degrees and was too cold".Root cause: while we are unable to determine the origin of the reported complaint, our observations reasonably suggest that it is not manufacturing related.Based on the information provided by the customer, the most likely root cause is failure to follow dfu, as the customer states that the temperature of the theatre was set at 14 degrees and was too cold.The dfu instructs that the iol and viscoelastic should be allowed to attain room temperature prior to use.The file states the use of a cartridge, dfu instructs that cartridge should be used at operating room temperatures between 18°c and 23°c.Not following the correct temperature may result in delivery issues and/or damage.All product and batch history records are quality reviewed prior to product release.Additional information provided in h.3., h.6.And h.10.The manufacturer internal reference number is: (b)(4).
|