A field service engineer (fse) spoke with the customer via the telephone.The customer confirmed that there were no obstructions from stopping the main arm from moving.The fse had the customer power down the analyzer and the pc at the same time, then turn them both on.Afterwards, the customer was able to perform daily check successfully.The customer completed start up, controls and patient samples with no system errors.The aia-2000 analyzer returned to normal operation.No further field action required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 15jun2019 to aware date 15jul2020.No other similar complaints were identified during the search period.The aia-2000 operator's manual states the following: [4211] main arm y-axis home detection failure cause: the home sensor failed to activate after the main arm y-axis moved toward the home position.If retry fails, the measurement result will be flagged (mf flag).Solution: contact tosoh service center or local representatives.The probable cause is attributed to dropped cups which caused a mechanical jam of the main arm.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|